7 Misunderstandings About HPLC and UHPLC Columns

Because HPLC has the advantages of high speed, high separation effect, high sensitivity, etc., it has wider and wider applications. HPLC chromatographic column is an important part of the HPLC system. It needs to be used carefully when used, and it must be maintained after use to avoid problems in the next use. Frequent replacement of the chromatographic column is very expensive, so in order to save costs, it is necessary to extend the service life of the chromatographic column to a greater extent. At the same time, the chromatographic column must be carefully maintained to maintain accuracy and consistency every time the result is obtained. HAWACH editor has summarized 7 misundertandings about the columns hoping it would extend its service life.

Misunderstanding 1: HPLC column cannot be backflushed

Under normal circumstances, the designed pressure resistance value of the liquid chromatographic column is much higher than the maximum operating pressure. When replacing the column, backflushing can clean some substances with strong adsorption properties at the head of the column, and we flush out these residual substances to prevent pressure increase. But if the chromatographic column you buy uses a large particle size packing at the column head, this part of the packing will be flushed out when you backflush. Therefore, you have to check the liquid chromatography column you are using before choosing whether to backflush.

Misunderstanding 2: All C18 columns are the same

In the early HPLC system, C18 was the bonded stationary phase for standard reverse chromatography, so C18 was used as the standard for reverse chromatography columns. However, the times will not remain stagnant. People’s pursuit and understanding of science has become more and more profound, more stationary phases have appeared, and many C18 chromatographic columns have been born. Although silica gel is also used as the matrix, each has its own specific filler bonding synthesis process, so the chromatographic performance is different. So simply because the names are all C18 columns, it is wrong to think that the performance is the same.

XChroma HPLC Columns High Purity HPLC Columns XChroma HPLC Column

Misunderstanding 3: Once air enters the chromatographic column, it will be damaged

We all know that when the column is not connected to the chromatograph, it is necessary to ensure that the column is tightly sealed. However, in practical applications, even if a small amount of air enters the end of the chromatographic column, it is not a serious matter. When the chromatographic column is connected to the chromatograph for use, the air will be squeezed out by the solvent in a short period of time during the initial pressurization of the system. Therefore, do not assume that the column is damaged just because a small amount of air has entered the column.

Misunderstanding 4: The guard column is unnecessary

In fact, there are many advantages to using a guard column. Various reagents are often used in the mobile phase of HPLC, and some insoluble or impure substances are mixed in it. Most insolubles can be removed by filtering the mobile phase with a sieve or membrane filter, but very fine particles cannot be completely filtered. In addition, after the mobile phase is modulated, there is a possibility that insoluble and impure materials may fall through the air, container, human body, etc. These insoluble and impure substances can contaminate the column, block the column, or have an adverse effect on the analysis.

At this time, by installing an analytical guard column in front of the chromatographic column, it can prevent its generation and protect the chromatographic column. Compared to replacing an expensive analytical column, adding or replacing a guard column requires very little expense. The HAWACH liquid chromatography guard column has almost no dead volume and is easy to replace. It is suitable for the high pressure of the UHPLC system and intercepts those strongly retained and non-adsorbable compounds. The filtration of sample and mobile phase can maintain the adsorption capacity of the guard column for chemical contaminants, maintain column efficiency for a longer time, and extend the service life of the chromatographic column.

Misunderstanding 5: Reversed-phase chromatographic column cannot use pure water phase

Some chromatographers experience phase collapse when using low organic solvent content or pure water as the mobile phase of reversed-phase chromatography columns. Therefore, some people think that reversed-phase chromatography columns cannot use pure water. But in fact, the reversed-phase chromatography columns sold on the market (such as polar embedding and polar capped columns) are all water-wettable, and their surface characteristics allow the use of pure water, which will not cause collapse or retention time shift.

Misunderstanding 6: The smaller the particle size and the higher the pressure of the column packing, the better the separation effect

The performance of a chromatographic column cannot be evaluated solely by whether the filler is ultra-small particle size and ultra-high column pressure. Modern researches on the column characteristics of chromatographic columns have become more and more in-depth, and some methods have been developed to improve the efficiency of chromatographic columns. For example, the column efficiency of the new surface porous material column is as good as that of the sub-2µm UHPLC column. It’s column pressure is very low, compared with the conventional column.

Misunderstanding 7: Column pressure will not affect the chromatographic separation effect

In recent years, the problem of column pressure has attracted the attention of many chromatographers. Many parameters of chromatogram are affected by column pressure, including the molar volume of some solutes, stagnation volume, column porosity, retention factor and mobile phase etc. Our attetnion will not be attracted, if operating the column at a pressure of 13.789MP, even with small difference in retention time. However, when the column pressure is close to 2000psi (13.789MPa), the effect of column pressure may be quite obvious.

Conclusion
In recent decades, high performance liquid chromatography has been widely used in the fields of life sciences, drug analysis, food analysis, environmental monitoring and other fields. People’s continuous research on liquid chromatography has also produced more liquid chromatography applications, solutions and methods, etc. It is believed that the application field of liquid chromatography will be broader in the future.